FDA Posts Class II Recall for Nerivio Infinity TENS Device for Migraine

08/04/2026

KEY TAKEAWAYS

  • The FDA posted a Class II recall for Nerivio Infinity, a distal transcutaneous electrical stimulator used for acute migraine treatment.
  • The recall involves a potential battery overheating issue during charging, which may result in smoke and odor.
  • Corrective actions include mobile app and firmware updates, with the firmware designed to disable the device if overheating is detected.

The Food and Drug Administration (FDA) posted a Class II recall for Nerivio Infinity (Theranica Bioelectronics, Netanya, Israel), a distal transcutaneous electrical stimulator (TENS) device for acute migraine treatment. The recall was initiated by the firm on March 23, 2026, and posted by the FDA on July 31, 2026. The recall status is open and classified.

Recall Details

  • Recall number: Z-2855-2026
  • Recall Event ID: 99307
  • 510(k) number: K250405
  • Affected product: Nerivio Infinity, Model/Catalog No. FGD000100
  • Affected software: Firmware version 4.1.2.0; app version 7.28.0.11
  • Quantity in commerce: 11,886 units

According to the FDA recall entry, the recall was initiated because of the potential for the device battery to experience a thermal runaway event during charging, which could result in smoke and odor. The FDA determined the cause to be component design or selection.

Theranica issued a stop shipment from distributors beginning March 22, 2026. The devices were released from hold on April 30, 2026, after a firmware update was released on April 19, 2026. Beginning May 13, 2026, customers received in-app messaging with instructions for corrective action.

Users were prompted to update both the mobile application used to control the device and the device firmware. The firmware update is intended to permanently disable the device if an overheating event is detected, while the software update adds enhanced warning functionality if a device malfunction is detected.

Source

US Food and Drug Administration. Class 2 Device Recall Nerivio Infinity. Recall No. Z-2855-2026. Posted July 31, 2026. Accessed August 4, 2026. https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?id=221248